Header image  
Report of the Eighth Census Conference on Newborn ICU Design
Committee to Establish Recommended Standards for Newborn ICU Design
Robert D. White, MD, Chair
January 26, 2012 Clearwater Beach, FL
 
   September 18, 2012
 
 
 
Standard 23 - Procedure Lighting in Infant Care Areas

Separate procedure lighting shall be mounted at each infant bed. The luminaire shall be capable of providing no less than 2000 lux at the plane of the infant bed, and must be framed so that no more than 2% of the light output of the luminaire extends beyond its illumination field. This lighting shall be adjustable so that lighting at less than maximal levels can be provided whenever possible.

Interpretation: Temporary increases in illumination necessary to evaluate a baby or to perform a procedure should be possible without increasing lighting levels for other babies in the same room.

Since intense light may be unpleasant and harmful to the developing retina, every effort should be made to prevent direct light from reaching the infant's eyes. Procedure lights with adjustable intensity, field size, and direction will help protect the infant's eyes from direct exposure and provide the best visual support to staff.

It is preferable that the procedure light be either mounted on the headwall or incubator in lieu of a floor stand. This will maximize the space around the infant work area and minimize trip hazards.


 
Next

 
Additional Information
 
 
   
last updated September 19, 2012  Kathleen Kolberg, University of Notre Dame